Biosen C-Line Clinic 5214-0051-2200
Results in 20-45 seconds
Imprecision < 1.5%
Up to 120 results an hour
Multi-lingual touch screen display
with step by step instructions
Sports Performance
Diabetes Care
Hemoglobin
& Hematocrit
Lab-accurate glucose and lactate
measurement from one sample
Hemolyzed sample stable for up
to five days
Long life chip-sensor technology
Download
PDF
Pack includes
Specifications
Methodology ................................................... Electrochemical
Models
Principle.......................................... Enzymatic-amperometric
Sample type....................... Whole blood, plasma or serum
Biosen C-Line
GP+
Up to 5 samples
Clinic
Up to 20 samples
Sample size...............................................................20 µL (10 µL)
Measuring range............................ Glucose 0.5–50 mmol/L
Lactate 0.5–40 mmol/L
Imprecision CV...........................................≤1.5 % (12 mmol/L)
Biosen S-Line
Lab+
Biosen S-Line Lab+
5222-0056-6200
Analyzer with multi-language software, EDP, printer
port, power adaptor, integrated barcode reader
(optional) and user manual. 100-240V A/C, 50-60 Hz.
CE marked
Play
Video
Biosen C-Line GP+
5213-0051-6200
Up to 63 samples
For sports training use only (USA).
Calibration................................................. Automatic, 3 modes
Point of Care Diabetes Care | 17
Central Laboratory
Stores up to 1,000 results
Maternal &
Women’s Health
Suitable for whole blood,