4.3 Performing the Measurement Cont.
7.
9.
11.
7. Place camera tip under patient’s tongue as you would a thermometer.
Observe the MOTION, INTENSITY AND FOCUS bars on the VIDEO QUALITY
screen. Have patient gently bite down on the lens cover to keep the camera
steady. This will be seen in the MOTION green bar. The moving curser should
remain within the green bar. Ask patient to tilt the chin slightly down and to
breathe through the nose so that the tongue remains still. In the INTENSITY
green bar, the moving curser should remain within the green bar. Move the
camera tip slowly around until you 昀椀nd an area with a high capillary density.
8. Adjust with the focus wheel on the camera, watching the cursor on the left
panel. When properly focused, the cursor will fall within the green bar. If you
have trouble achieving proper focus, gently move camera tip to a different
area under the tongue. (Light intensity is adjusted automatically when using
the KK Technology camera).
9. When the focus is set, the video camera will start recording. You will notice
the bar 昀椀lling upwards in green on the right panel. When the progress bar is
fully green, the measurement is complete.