USER'S MANUAL
EVOLUTION VETERINARY DENTAL UNITS
22 � EN
ERROR MESSAGES
DISPLAY
DESCRIPTION OF ERROR
SOLUTIONS
Error 1
Overheating / overloading of electronics
> ➊ Disconnect device from power supply
➋ Wait 5 minutes and allow the system to cool
➌ Switch the device back on again and restart
the function
Error 2
Pedal of the foot control pressed during
switching on
> Do not press the pedal of the foot control
Error 4
Display keys pressed during switching on
> Do not press the display keys
Error 5
Running time limiter as a result of
15 minutes of continuous operation
> Check the pedal of the foot control
> Do not press the pedal of the foot control any more
[release completely]
Error 3
Error 6
Error with »Electric motor /
Endo contra-angle handpiece /
Piezo scaler«
> ➋ electric motor / Endo contra-angle handpiece /
replace tip
> Switch the power supply off and on again
> Restart system
Reboot
Error
> ➊ Check the electric motor / Endo contra-angle
handpiece / scaler is correctly attached to the
coupling
e.g.: 05 6303
> Contact an authorised iM3 service partner
immediately
If the error messages described cannot be resolved
a check by an authorised service partner is required.
For a list of iM3 Authorised Service Partners in your country
please visit:
Europe & UK: www.im3vet.com
USA: www.im3vet.com
Australia/Asia: www.im3vet.com.au