INDUSTRIAL SECTOR CAT 2024 - Flipbook - Page 16
PG030 - Stainless steel / brass 昀椀llable pressure gauges
Product features:
• Waterproof and corrosion resistant, stainless steel
case and brass connection
• Glycerine 昀椀lling available
• Front or back 昀氀ange available for panel mounting
• Brannan quality and accuracy
• Personalised instrumentation available on request
• For outdoor and high humidity installations
• Wide choice of pressure ranges and dial diameters
• Wide range of associated accessories available
Pressure
Applications:
•
•
•
•
•
•
•
•
•
Pneumatics
Hydraulics
Heating, ventilation, air-conditioning
Food and beverage industry
Pulp and paper industry
Machine and apparatus construction
Pumps, compressors, hydraulic presses
Outdoor installations
High humidity installations
Product speci昀椀cations:
Dial diameters
Design
Safety category
Accuracy class
Ranges
Case
Process connection
Connection orientation
Connection material
Bezel
Measurement for life.
Measuring principle
Movement
Window
Pointer
Filling
Working pressure
Operating temperature
Storage temperature
Temperature e昀昀ect
Protection
Calibration
16
40mm, 50mm, 63mm, 80mm, 100mm & 115mm
BS EN 837
100mm & 115mm: S1 (blow-out device gauge). Others: 0 (no blow-out device)
40mm: 2.5%, 50mm, 63mm, 80mm & 100mm: 1.6%, 115mm: 1%
-1 to 9 bar & psi through to 0 to 1000 bar & psi
Stainless steel 304
1/8”, 1/4”, 3/8” & 1/2” BSP/NPT/BSPT (others available on request)
40, 50, 60 & 80mm - bottom & centre back
100mm - bottom & lower back, 115mm - bottom
Brass
Stainless steel 304. 50mm & 63mm dia dials: rolled bezel,
100mm dia dial: 昀氀at bezel, 115mm: bayone bezel
Bourdon tube Cu-Alloy
Cu-Alloy
40mm, 50mm, 63mm, 80mm & 100mm: polycarbonate (tempered glass optional)
115mm: tempered glass
Black aluminium
Dry or glycerine 昀椀lled
Steady:
75% full scale range
Fluctuating: 60% full scale range
Short time: full scale range
Ambient:
-20 to +60°C
Medium:
-20 to +60°C
-40 to +60°C
When temperature of the pressure element deviates from reference temperature of
20°C the maximum additional error is +/-0.4 per 10°C of the span rising and falling
IP 65
UKAS certi昀椀cation, certi昀椀cate traceable to recognised national standard
or statement of conformity, if required
brannan.co.uk