USER'S MANUAL
EVOLUTION VETERINARY DENTAL UNITS
OPERATING & TRANSPORT CONDITIONS
CONDITION
Temperature
Relative humidity
Air Pressure
Installation Altitude
OPERATION
+10°C to + 35°C
15% to 80%
800 hPa to 1060 hPa
Max 2,000 metres above sea level
TRANSPORT & STORAGE
- 20°C to + 70°C
8% to 80%
700 hPa to1060 hPa
INSTALL REQUIREMENTS
ROOM
Width
Length
Height
REQUIREMENT WITH
TABLE
200cm
300cm
> 200cm
ELECTRICAL
Mains Voltage
Mains Frequency
Primary Fuse
WATER
Type
Filter
Extras
REQUIREMENTS
110V or 240V (at Option) Non switchable
50/60Hz
10A(110V) 5A (240V) Circuit Breaker
REQUIREMENTS
Distilled or Ionized water only to be used
iM3 Straw fitted to water bottle, (replaced yearly)
iM3 CLS present in CLS bottle or iM3 straw fitted, (replaced yearly)
The Evolution Dental unit must be installed in accordance
with the instructions/unboxing video or by an iM3 Authorized
Service Provider.
Wall mounted version needs to be installed only by an
iM3 Authorised Service Provider.
WARNING: To avoid risk of electrical shock, this equipment
must be connected to a supply with protective earth.
To avoid the risk of electric shock always switch off the power
to the unit in the electrical panel before opening or touching
internal components.
Unboxing instructions for the Evolution range can be
downloaded from www.im3vet.com
9 � EN