Micronic Product Catalog 2023 - Online version - Flipbook - Page 8
I N N O V A T I O N
IN
SAMPLE
STORAGE
0.30 ML
0.40 ML
0.75 ML
0.80 ML
Page
20
Page
21
Page
22
Page
23
1.10 ML
Page
24
1.40 ML
Page
25
Ultra low temperature rack
Optional (DM = 2D Data-Matrix code on
rack bottom)
Micronic 96-1
DM
Micronic 96-1
DM
Micronic 96-2
DM
Micronic 96-3
DM
Micronic 96-3
DM
Micronic 96-4
DM
Thread type
Tube type
Inner tube shape
External
Screw Cap tube
V / Vr
External
Screw Cap tube
V
External
Screw Cap tube
V
External
Screw Cap tube
V
External
Screw Cap tube
V
External
Screw Cap tube
V
Working volume with Push Cap (+21ºC)*
Working volume with Screw Cap (+21ºC)*
Working volume with Screw Cap (-20ºC)*
Working volume with Screw Cap (-80ºC)*
0.25ml / 0.20ml
0.32ml / 0.23ml
0.25ml / 0.23ml
0.25ml / 0.23ml
0.38ml
0.40ml
0.38ml
0.36ml
0.57ml
0.60ml
0.55ml
0.55ml
0.68ml
0.75ml
0.65ml
0.65ml
0.85ml
0.93ml
0.87ml
0.77ml
1.03ml
1.11ml
1.05ml
0.94ml
Yes
No
No
Yes
Yes
No
No
Yes
Yes
No
Yes
Yes
Yes
No
No
Yes
Yes
No
No
Yes
Yes
No
Yes
Yes
Screw Cap (IG)
Push Cap
Screw Cap (IG)
Push Cap
Screw Cap (IG)
Push Cap
Screw Cap (IG)
Push Cap
Screw Cap (IG)
Push Cap
Screw Cap (IG)
Push Cap
Outer diameter
Tube height
Tube height with Push Cap
Tube height with Screw Cap IG
Tube height with Screw Cap
8.8mm
16.4mm
17.8mm
18.7mm
19.9mm
8.8mm
21.8mm
23.2mm
24.1mm
25.3mm
9.0mm
30.0mm
31.4mm
32.3mm
33.5mm
8.8mm
34.8mm
36.2mm
37.0mm
38.3mm
9.0mm
41.0mm
42.7mm
43.6mm
44.6mm
9.0mm
48.0mm
49.4mm
50.3mm
51.5mm
Available precapped
Twist-lock
Snap tube option
Sterilization option
DNase / RNase free
Class 7 clean room production
Minimum storage temperature**
Availability
Yes
Yes
No
Yes
Yes
Yes
Vapor phase LN2
Bulk, Rack
Yes
Yes
No
Yes
Yes
Yes
Vapor phase LN2
Bulk, Rack
Yes
Yes
No
Yes
Yes
Yes
Vapor phase LN2
Bulk, Rack
Yes
Yes
No
Yes
Yes
Yes
Vapor phase LN2
Bulk, Rack
Yes
Yes
No
Yes
Yes
Yes
Vapor phase LN2
Bulk, Rack
Yes
Yes
No
Yes
Yes
Yes
Vapor phase LN2
Bulk, Rack
2D Data-Matrix coded (bottom)
Alphanumeric coded (bottom)
Non-coded (bottom)
Human readable coded (bottom)
Capping options
*
**
If more sample volume is used in a tube than the specified working volume, it could be possible that deformation of the tube occurs. Request Test Report TR023201 for more information.
Temperature range of the material of the tubes, caps and racks is approved from -80ºC to +100ºC. In practice tubes are used in lower temperatures until vapor phase LN2
(-150ºC to -196ºC). Micronic does not recommend to store tubes in liquid phase nitrogen (LN2). In practice tubes are only snap freezed in LN2 (without submerging the cap).
*** For -80ºC storage of the 2.00ml tube its recommended to always use a rack with cover.
8
MICRONIC | Innovation in sample storage