USER'S MANUAL
54 � EN
EVOLUTION VETERINARY DENTAL UNITS
ADJUSTING REGULATOR AIR PRESSURE
With the Dental Unit pressurised and air lines turned ON,
the black knob on top of the regulator is pulled up from the
locked to the free position to adjust air pressure (see picture
page 21). Depress the black knob after setting to lock
pressure adjustment. The regulator gauge registers working
or air line pressure which may be increased by turning
clockwise and lowered by turning anti clockwise. The
pressure air line gauge will register the working pressure,
which should be 70 psi./ 6BAR. Please note the regulator
pressure us preset by uM3 before delivery of your unit
CIRCUIT BREAKER PUSH TO RESET
Unit starts automatically. Isolate power
supply and relieve air pressure by opening
tank drain valve prior to commencing non
electrical service tasks
Danger High voltage used on this equipment.
Isolate from mains before removing any covers,
which should only be carried out by a qualified
electrician . Ensure power outlet is adequately
rated for this unit.
COMPRESSOR MAINTENANCE
FEATURE
DAILY
Compressor
Regulator Auto
Drain Filter
Regulator Air
Pressure
WEEKLY
MONTHLY
YEARLY
Turn off power and unplug from
wall. Drain moisture from main
compressor tank by opening
the tank drain on the top of the
white tank marked with a yellow
sticker. Catch the fluid from the
white drain tube into a container.
Keep the drain open until all air is
removed from the tank.
Automatic when
tank drained of
moisture
Check that correct
air pressure is
being delivered
For a yearly service kit or to arrange for a yearly service of your dental unit please contact
support@im3dental.com or www.im3vet.com
Replace regulator
filter