USER'S MANUAL
EVOLUTION VETERINARY DENTAL UNITS
HIGH SPEED HANDPIECE
Handpieces should be thoroughly cleaned to remove
any foreign particles, eg ; blood, saliva . Make sure that
the turbine cartridge and head cavity are clean before
autoclaving. If autoclaved with bloody particles in the
turbine housing, autoclaving will cause particles to be
scorched and damage to the turbine may result.
The exterior of the handpiece may be cleaned with
any good surgical disinfectant. Be sure to rinse and dry
thoroughly. Lubricate handpiece turbine with MD-30
Lubricant. Do not submerge the handpiece in liquid. Refer
to handpiece operating manual.
Lubricate handpieces using MD-30 Lubricant
after cooling down, do not allow handpiece to
sit in sealing bag overnight or for a long period.
Immediately operate handpiece as described in
the following step.
4.
After lubricating items, insert bur into handpieces and
operate for a few seconds to remove excess lubricant.
Do not autoclave the high speed handpiece
with a bur in the chuck.
ULTRASONIC SCALER
For cleaning information read in full the
sub-manual for the Scaler
LOW SPEED HANDPIECE
Prepare handpiece and attachments for sterilization by
carefully cleaning the exterior surface. Thoroughly scrub,
rinse and clean away any residual solution and particles.
Remove moisture with a towel or use air from the 3 way air
water syringe.
Prior to sterilization, lubricate the handpiece using MD-30
Lubricant. Do not submerge the handpiece in liquid
HANDPIECES: Apply oil through the handpiece air inlet
tube. As pictured on the iM3 MD-30 bottle
ATTACHMENT: Remove attachment from handpiece and oil
through both end openings.
After lubricating items, assemble and operate
for a few seconds to expel excess lubricant
1.
Seal items in sterilization tubing bags. Make sure to
remove bur from handpiece prior to sealing.
2.
Place handpiece or attachments in sterilization trays of
dry heat sterilizer or steam autoclave. Do not exceed
temperatures of 275˚F or 135˚C. Sterilize according to
manufacturer’s instructions.
3.
After sterilization cycle is completed, allow handpieces
and attachments to cool down then lubricate
handpiece and attachments thoroughly. Make sure
to use a different container after sterilization cycle, to
prevent re-contamination.
47 � EN
Fig.44 - Handpiece Bridge
Fig.45 - Prior to sterilization lubricate the handpiece using iM3's MD-30