USER'S MANUAL
EVOLUTION VETERINARY DENTAL UNITS
46 � EN
INFECTION CONTROL OF
OTHER INSTRUMENTS
For turbines, ultra sonic scalers and other instruments
please refer to their user manuals supplied with the units.
DISINFECTION OF UNIT WATER
LINES
The iM3 Straw should be used at all times on
the water line of your dental units. As standard
your unit will have come with a Straw, please
follow the instructions carefully on how to use
this item.
It is a violation of federal/local law to use this
product in a manner inconsistent with its
labeling.
INITIAL INSTALLATION:
Unscrew the water bottle on your dental unit.
» Fill bottle with distilled quality water.
» Fill the bottle, leaving an air space of at least 1”
from the top of the bottle
(do not over fill the bottle - adequate airflow is
necessary to ensure proper pressurization).
To install Luer Lock Straw
1.
To Detach Straw
2.
Insert barbed end of fitting
into the precut pickup tube.
3.
Insert wide end of fitting
into Straw.
4.
Turn the Straw right to
connect. Hand tighten until
connection is snug.
Turn to the left to remove
Fig.43 - Installation of water bottle straw
The iM3 Straw includes a Luer Lock.
Proper installation of the luer lock is shown in the
diagrams. It is important to cut the existing pickup tube
to the proper length so that the bottom of the iM3 Straw
is at least a 1/4” from the bottom of the bottle.
INITIAL TREATMENT: iM3 Straws are engineered to
automatically produce a “Shock” treatment after initial
connection and treatment of dental water. Slowly purge the
waterlines attached to the dental unit into a cup until you
see the light blue colour. The colorant is an indicator the
Shock product is in the line(s). Allow lines to remain unused
for a minimum of 12 hours. For best results, allow Shock
to remain in the line(s) over the weekend. Prior to use of
dental unit, flush dental lines until blue colour disappears
(approximately 4 ounces).
MAINTENANCE TREATMENT: Fill water bottle with
distilled water. The dental water is automatically treated
as it runs through the Straw Cartridge. Replace the Straw
Cartridge per the Straw Capacity & Replacement Schedule,
at least yearly.
Fig.42 - Evolve Veterinary Dental Unit Disinfect all elements
HAZARDS TO HUMANS AND DOMESTIC
ANIMALS - CAUTION: In the event of a
rupture of the cartridge, avoid contact
with the silver ion exchange resin which
may cause eye and skin irritation. Avoid
direct contact with eyes, skin and clothing.
Wash thoroughly with soap and water
after handling and before eating, drinking,
chewing gum, using tobacco or using the
toilet. Remove and wash contaminated
clothing before reuse. Place debris into a
sealed plastic bag and discard in trash.
ENVIRONMENTAL HAZARDS: This pesticide
is toxic to fish and aquatic invertebrates.
FIRST AID: If in eyes: Hold eye open and
rinse slowly and gently with water for
15–20 minutes. Remove contact lenses
(if present) after the first 5 minutes
then continue rinsing. If on skin: Take
off contaminated clothing. Rinse skin
immediately with plenty of water for 15–20
minutes. Call a poison control centre or
doctor for treatment advice.
Handpiece
Refer for more details to the handpiece
operating manual.