EVOLUTION VETERINARY DENTAL UNITS
USER'S MANUAL
9. TIPS FOR OPERATORS AND SERVICE TECHNICIANS
Any repairs exceeding usual routine maintenance must
be carried out by suitably qualified personnel.
WARNING
Infection due to contaminated unit
Contaminated units can cause infections.
» Clean and disinfect the iM3 Bottle System before working on the unit.
» Wear suitable personal protective equipment when working
(e.g. impermeable gloves, protective goggles and mouth and nose protection).
ERROR
Low water flow rate
No water flow
POSSIBLE CAUSE
SOLUTION
Fine filter blocked in the riser pipe
Replace the fine filter.
Blockage in the riser pipe
Remove and clean riser pipe.
Blockage in the PUR hose
Remove and clean the PUR hose, replace if
necessary.
Kink in the PUR hose
Reposition the PUR hose.
Blockage in the riser pipe
Remove and replace riser pipe.
Blockage in the PUR hose
Remove and replace the PUR hose.
No pressure
Switch on the compressed air supply.
Screw in the bottle as far as it will go, to open
the stop valve.
Fluids leakage from the
bottle system
PUR hose is not attached correctly
Reconnect the PUR hose with a hose safety
device.
Air blows off audibly
Bottle not completely screwed in
Turn bottle to screw in completely.
Safety valve blows off
Pull the ring of the safety valve to vent.
10. MANUFACTURER OF BOTTLES
Manufactured and serviced by:
iM3 Dental Limited
Unit 29, Duleek Business Park,
Duleek, Co. Meath,
A92 N72W, Ireland.
Phone: +353 (0) 1 691 1277
Email: sales@im3dental.com |
support@im3dental.com
Website: www.im3vet.com
iM3 Inc.
12414 NE 95th St., Vancouver, WA,
98682, USA
Phone: +1 800 664 6348
Fax: +1 360 254 2940
Technical Support: +1 800 664 6348
Email: info@im3usa.com
Website: www.im3vet.com
iM3 Pty Ltd.,
21 Chaplin Drive, Lane Cove, NSW
2066, Australia.
Phone: +61 2 9420 5766
Fax: +61 2 9420 5677
Email: sales@im3vet.com
Website: www.im3vet.com.au