EVOLUTION VETERINARY DENTAL UNITS
6. DISINFECTION AND CLEANING
WARNING
Health hazard due to contaminated unit
Contaminated units can cause infections.
» Clean and disinfect the iM3 Bottle System
before working on the unit.
» Wear suitable personal protective equipment
when working (e.g. impermeable gloves,
protective goggles and mouth and nose
protection).
38 � EN
6.1 Daily after the end of treatment
The bottle contents must be changed every day,
otherwise there is a risk of contamination.
The following are required for cleaning:
» Non-foaming disinfectant/cleaning agent that is
compatible with the materials.
» Clean the bottle with hot water (≤ 90 °C) and a cleaning
brush.
Cleaning and disinfecting surfaces
NOTICE
Device malfunction or damage due to incorrect
cleaning and disinfecting agents
Guarantee claims may become invalid as a result.
» Do not use any foaming agents, e.g. household
cleaning agents or instrument disinfectants.
» Do not use abrasive cleaners.
» Do not use agents containing chlorine.
» Do not use any solvents like acetone.
NOTICE
Device malfunction or damage due to damaged/
worn bottles
Guarantee claims may become invalid as a result.
The bottles can burst.
Bottles are subject to operational wear and tear
and have a limited service life*.
» Exchange bottle after one year or 50 automatic
treatments.
» Damaged, scratched or opaque bottles must
be replaced immediately.
* The expiry date is lasered on the bottom of the bottle,
along with the corresponding pictogram.
Fig.38 -Bottles Expiry date on bottle
» Regularly clean and disinfect surface and thread of the
bottle and bottle receptacle, e.g. with FD 366 sensitive
top wipes.
6.2 Once or twice a week/break-time
preservation
The following are required for cleaning:
» Non-foaming disinfectant/cleaning agent that is
compatible with the materials.
» Rinse the bottle with hot water.
» Clean the bottle with a cleaning brush or in the
washer-disinfector.