USER'S MANUAL
28 � EN
EVOLUTION VETERINARY DENTAL UNITS
DISTILLED WATER BOTTLE & CLS BOTTLE
CLS
WATER
Fig. 26 - Water & CLS Bottles
www.im3vet.com/manuals
Fig. 27- Side of Unit controls
Handpiece Pressure
CLS & WATER
SWITCH
HANDPIECE WATER
ADJUSTMENT
Your Evolution unit is supplied with a sterilising straw and a bottle of CLS,
please follow the instructions for use before operating the dental unit.
Watch the unboxing video supplied for more information on CLS and the
straw, or visit www.im3vet.com for further information.
The water bottle is marked with a water droplet and the CLS bottle is
marked with the words CLS. The CLS bottle is UV protected and a normal
water bottle should not be used for CLS as it is light sensitive.
The water bottles will automatically pressurise when the bottles are
attached to the dental unit and depressurise when they are detached from
the unit.
Only tighten bottles hand tight, over tightening the bottles will reduce the
life of the bottles and damage the bottle holder thread.
Only distilled/ionised water should be mixed with the CLS, if
normal/tap water is used in conjunction with CLS then the unit can be
blocked. Always follow the instructions on the CLS bottle for dilution
amount.
Only Distilled/ionised water should be used in the water bottle, if normal/
tap water is used in conjunction with the straw then the unit can be blocked.
If CLS is not used in the CLS bottle then the filter should be detached and a
iM3 straw should be attached. If no straw is used then the machine will not