Combined-languages-d1 - Flipbook - Page 26
USER'S MANUAL
EVOLUTION VETERINARY DENTAL UNITS
OVERVIEW OF DISPLAY
26 � EN
CHANGING / SAVING THE POWER ON
THE ULTRASONIC SCALER
Factory Settings
Program 1 Display shows P1: 30
Program 2 Display shows P2: 20
The power can be set at P1, P2, P3
for a minimum of 1 to maximum 40
Program 3 Display shows P3: 15
PLUS key: Increase power, maximum 40,
possible during use
MINUS key: Reduce power, minimum 1,
possible during use
*By holding PLUS/MINUS key the values are continuously
increased / decreased.*
Step 1: Press the program key (P1,P2 or P3).
Step 2: Increase power
Step 3: Reduce power
Step 4: Press program key P (P1, P3 or P3) for
approximately 2 seconds to save.
Program 1
40,000
rpm
The Set values and an acoustic signal is emitted
to confirm that all the settings have been
saved.
Program Key
Ncm
Ncm
Fig. 21 - Display Screen Ultrasonic Scaler
ERROR MESSAGES FOR SCALER IF USING THE DISPLAY
DISPLAY
DESCRIPTION OF ERROR
SOLUTIONS
Error 1
Overheating / overloading of electronics
> ➊ Disconnect device from power supply
➋ Wait 5 minutes and allow the system to cool
➌ Restart the function
Error 2
Pedal of the foot control pressed during
switching on
> Do not press the pedal of the foot control
Error 4
Display keys pressed during switching on
> Do not press the display keys
Error 5
Running time limiter as a result of
15 minutes of continuous operation
> Check the pedal of the foot control
> Do not press the pedal of the foot control any more
[release completely]
Error 6
Error with »Piezo Scaler«
> ➊ Check that the Piezo Scaler is correctly attached to
the coupling
> ➋ Replace the tips
> Switch the power supply off and on again
> Restart system
Reboot
Error
e.g.: 05 6303
> Immediately contact an authorised
iM3 service partner(see pg6)