TECHNICAL DATA
MODEL
D-2.0 E
5 V DC ±10%, from W&H controller
V (supply voltage of the device)
Power supply:
Power consumption:
Dimensions in mm
[WxDxH]:
Weight:
1.5 VA
155.5 x 29 x 122
280 g
A M B I E N T CO N D I T I O N S
Temperature during storage and transport -40°C to +70°C [-40°F to +158°F]
Humidity during storage and transport 8% to 80% [relative], non-condensing
Temperature during operation +10°C to +35°C [+50°F to +95°F]
Humidity during operation 15% to 80% [relative], non-condensing
Permitted ambient pressure 70 – 106 kPa
Pollution level 2
Overvoltage category II
Altitude up to 3,000m above sea level
Classification according to Paragraph 6 of the General Specifications for the Safety of
Medical Electrical Equipment according to IEC 60601-1 / ANSI/AAMI ES 60601-1